Journal of Neurosurgical Sciences 2021 June;65(3):348-53
DOI: 10.23736/S0390-5616.20.04957-7
Copyright © 2020 EDIZIONI MINERVA MEDICA
lingua: Inglese
A new tool to improve pedicle screw placement accuracy in navigated spine surgery: a monocentric study
Giovanni NICOLETTI 1, Massimo FURNARI 1, Massimiliano GIUFFRIDA 1, Giancarlo PONZO 1, Domenico G. IACOPINO 2, Giacomo CAMMARATA 2, Gianluca SCALIA 1, 3, Francesca GRAZIANO 2 ✉
1 Division of Neurosurgery, ARNAS Garibaldi, Catania, Italy; 2 Department of Experimental Biomedicine and Clinical Neurosciences, School of Medicine, AOUP Paolo Giaccone, Palermo, Italy; 3 Division of Neurosurgery, Department of Biomedical and Dental Sciences and Morpho-Functional Imaging, University of Messina, Messina, Italy
BACKGROUND: Navigated instrumented spine surgery is burden by a low but significant screw mispositioning risks, respectively, for the 2D imaging system from 15% to 40% and, for the 3D imaging system, ranging from 4.1% to 11.5%. The primary objective of this study was to demonstrate the efficacy of a new screw-like tool in order to further decrease pedicle screws mispositioning rate during vertebral navigated spine surgery.
METHODS: Between January and June 2019 an initial case series of 18 patients were enrolled. All patients underwent a pedicle screw fixation, both in thoracic (Th10-Th12) and lumbosacral (L1-S1) spine, using O-arm (Medtronic Navigation, Louisville, CO, USA) and StealthStation S7 surgical navigation system (Medtronic Navigation). Evaluation of pedicle screws placement accuracy using the new tool in adjunct to the classic reference frame, was performed following the Gertzbein and Robbins classification.
RESULTS: A total of 94 screws have been placed. Among them, 98.9% were completely inside cortical bone (grade A) and only 1.1% with a breach of less than 2 mm (grade B).
CONCLUSIONS: Our new “screw-like” tool coupled to the classic reference frame device could improve accuracy during navigated spine surgery and potentially reducing to zero the risks for screw mispositioning.
KEY WORDS: Neuronavigation; Equipment and supplies; Neurosurgical procedures; Pedicle screws