Minerva Anestesiologica 2020 November;86(11):1143-50

DOI: 10.23736/S0375-9393.20.14422-5

Copyright © 2020 EDIZIONI MINERVA MEDICA

lingua: Inglese

A randomized trial comparing the Ambu® Aura-i™ and the Ambu® Aura Gain™ laryngeal mask as conduit for tracheal intubation in children

Lukas GASTEIGER 1 , Elgar OSWALD 1, Maya KEPLINGER 1, Gabriel PUTZER 1, Markus LUGER 1, Sabrina NEURURER 2, Christian KELLER 3, Berthold MOSER 4

1 Department of Anesthesiology and Intensive Care Medicine, Medical University of Innsbruck, Innsbruck, Austria; 2 Department of Medical Statistics, Informatics and Health Economics, Medical University of Innsbruck, Innsbruck, Austria; 3 Department of Anesthesiology, Schulthess Clinic, Zurich, Switzerland; 4 Department of Anesthesiology and Intensive Care Medicine, Spital Limmattal, Schlieren, Switzerland



BACKGROUND: The Ambu Aura Gain is a newer second-generation supraglottic airway device designed for fibreoptic bronchoscopy (FOB)-guided tracheal intubation.
METHODS: 57 patients between 18 months and six years of age were randomized to receive either the Ambu Aura-I (N.=29) or the Ambu Aura Gain (N.=28). Primary endpoint was the time for intubation. Secondary endpoints were the time and number of attempts for device insertion, the feasibility of FOB-guided intubation, the oropharyngeal leak pressure (OLP) the fibreoptic grade of laryngeal view and possible complications.
RESULTS: No difference was found in the time for intubation, the time for device insertion or the fibreoptic grade of laryngeal view. First-attempt device insertion was successful in all (N.=28) patients with Aura Gain (100%) and in 27 (97%) with Aura-i. In the Aura-i group one insertion failed. A significant difference in successful intubation was seen between the Aura-i and the Aura Gain (79% vs. 100%, respectively, P=0.0011). Also found was a significant difference in the mean OLP (SD) between the Ambu Aura-i and the Ambu Aura Gain (18 [3] vs. 20 [3] cmH2O, respectively; mean difference [MD] 2 cmH2O; P=0.005).
CONCLUSIONS: The Ambu Aura Gain served as a reliable device for FOB-guided tracheal intubation. Even if the time for intubation, when intubation was possible did not differ, the Aura-i showed only 79% intubation success, making it a doubtful device for FOB-guided tracheal intubation in cases of emergency and severe hypoxemia in small children.


KEY WORDS: Laryngeal masks; Biological oxygen demand analysis; Hypoxia