CASE REPORT
Minerva Chirurgica 2001 June;56(3):303-6
Copyright © 2001 EDIZIONI MINERVA MEDICA
language: Italian
Post-traumatic fracture and migration in the pulmonar artery of the catheter of a totally implantable venous access device. An unusual complication
Iannelli A., Kianmanesh R., Msika S., Marano A., Levesque M., Grandjean M., Hay J. M.
The case of a 59 year-old patient, who sustained a post-traumatic fracture of the silastic catheter of his totally implantable venous access device that migrated in the right pulmonary artery, is reported. The venous device was placed six months earlier for the treatment of metastatic spread of a primary unknown adenocarcinoma. The venous device was placed on the left side in consideration of a recent right supraclavicular node biopsy. The catheter was introduced through the left internal jugular vein and its peripheral end was positioned subcutaneously across the clavicle to be connected to the port chamber placed in the infraclavicular region. The accidental fracture of the catheter was attributed to a closed trauma occurred during the transport of a refrigerator on the homolateral shoulder. Treatment involved extraction of the migrated fragment through a percutaneous transfemoral angioradiological procedure. A few days later the chamber was removed and a new totally implantable venous access device was placed on the other side.