Minerva Cardioangiologica 2012 October;60(5):507-15
Copyright © 2012 EDIZIONI MINERVA MEDICA
language: English
From bench to bedside: initial experience with the Primus drug-coated balloon catheter
Briguori C. 1, Virmani R. 2, Kolodgie F. 2, Byrne R. A. 3, Steigerwald K. 3, Orlowski M. 4, Joner M. 3 ✉
1 Laboratory of Interventional Cardiology and Department of Cardiology, Clinica Mediterranea, Naples, Italy; 2 CVPath Institute, Inc., Gaithersburg, MD, USA; 3 German Heart Center and Medical Clinic 1, Klinikum rechts der Isar, Technical University, Munich, Germany; 4 Cardionovum GmbH, Bonn, Germany
AIM:Drug-coated balloon (DCB) technology has emerged as a promising therapy particularly in the treatment of coronary in-stent restenosis. Although a variety of devices are available for clinical use, clinical outcomes have been variable and scope for significant improvement exists.
METHODS: In a preclinical study, a total of 10 juvenile healthy farm pigs underwent catheter-based DCB deployment in coronary arteries with angiographic and pathological follow-up at 7 or 28 days.
Animals were randomly allocated to the PRIMUS or Dior® DCB (N.=10 per group) and evaluated by histopathology and morphometric analysis. In a first-in-man clinical study a total of 19 consecutive patients presenting with restenosis within drug-eluting stents were treated with the PRIMUS DCB. Clinical follow-up was performed out to 6 months.
RESULTS:Neointimal thickness was similar between the PRIMUS and Dior® DCB groups, while fibrin deposition and inflammation were more sustained in the PRIMUS group at 28 days.
In 19 consecutive patients presenting with in-stent restenosis of drug-eluting stents, treatment with the PRIMUS DCB catheter resulted in high procedural efficacy. There were no adverse clinical events observed out to 6 months.
CONCLUSION:The PRIMUS DCB demonstrates high preclinical safety and excellent acute performance and safety. Further studies are needed to delineate the relative merits of this novel DCB compared to other devices.