JOURNAL TOOLS |
Publishing options |
eTOC |
To subscribe |
Submit an article |
Recommend to your librarian |
ARTICLE TOOLS |
Publication history |
Reprints |
Permissions |
Cite this article as |
Share |


YOUR ACCOUNT
YOUR ORDERS
SHOPPING BASKET
Items: 0
Total amount: € 0,00
HOW TO ORDER
YOUR SUBSCRIPTIONS
YOUR ARTICLES
YOUR EBOOKS
COUPON
ACCESSIBILITY
ORIGINAL ARTICLE CARDIAC SECTION
The Journal of Cardiovascular Surgery 2019 April;60(2):259-67
DOI: 10.23736/S0021-9509.18.10690-2
Copyright © 2018 EDIZIONI MINERVA MEDICA
language: English
Modified sizing technique with newly designed tools to facilitate the valve sparing aortic root replacement “David” procedure with mid-term results
Sharaf-Eldin SHEHADA 1 ✉, Jaroslav BENEDIK 2, Maria SERRANO 1, Juri LURBASKI 1, Ender DEMIRCIOGLU 1, Daniel WENDT 1, Heinz JAKOB 1, Fanar MOURAD 1
1 Department of Thoracic and Cardiovascular Surgery, West German Heart and Vascular Centre Essen, University Hospital Essen, Essen, Germany; 2 Department of Cardiovascular Surgery, Helios Clinic Krefeld, Krefeld, Germany
BACKGROUND: Valve sparing root replacement differs in specific points. The main target remains to achieve a perfect intraoperative result and long-term stability. We aimed in this study to present our modified sizing technique for valve-sparing “David” procedure and its mid-term results.
METHODS: We present a retrospective single-center study. A newly designed sizing ring in addition to triple-armed forceps (Trifeet®) was used to measure the proper size of the Valsalva® prosthesis for patients undergoing David-procedure. Primary endpoints are intraoperative aortic regurgitation (AR) and early postoperative outcomes. Secondary endpoints included freedom from aortic regurgitation or reoperation and overall mortality.
RESULTS: A total of 63 consecutive patients who underwent David procedure between 09/2012 and 12/2016 were evaluated. Mean age was 52±15 years and 76.2% were male. Moderate to severe aortic regurgitation was reported in 60 (95.2%) patients. Four (6.3%) patients presented with type-A aortic dissection, 20 (31.7%) patients had bicuspid and 3 (4.8%) had a unicuspid aortic valve, 2 (3.2%) patients had a prior aortic valve repair. Intraoperative echocardiography revealed no 34 (54%), trace 26 (41.2%) or moderate 3 (4.8%) AR. Stroke, myocardial infarction, and 30-day mortality occurred in 1 patient (1.6%). During follow-up 5 (7.9%) patients needed reoperation due to recurrent AR within a mean of 35±18 months. One could be re-repaired, and the other four underwent aortic valve replacement. A second patient died in the late follow-up.
CONCLUSIONS: Our modified sizing technique simplifies the “David-procedure” and allows to achieve a good intraoperative and mid-term results. However, these results have to be confirmed in a larger cohort with a long-term follow-up.
KEY WORDS: Heart valve prosthesis implantation - Follow-up studies - Cardiac surgical procedures