Home > Journals > The Journal of Cardiovascular Surgery > Past Issues > The Journal of Cardiovascular Surgery 2010 August;51(4) > The Journal of Cardiovascular Surgery 2010 August;51(4):461-6

CURRENT ISSUE
 

JOURNAL TOOLS

eTOC
To subscribe
Submit an article
Recommend to your librarian
 

ARTICLE TOOLS

Reprints
Permissions

 

ORIGINAL ARTICLES  NEW RESULTS IN THE ENDOVASCULAR MANAGEMENT OF ABDOMINAL AORTIC ANEURYSMS 

The Journal of Cardiovascular Surgery 2010 August;51(4):461-6

Copyright © 2010 EDIZIONI MINERVA MEDICA

language: English

Use of the Aorfix™ stent graft in patients with tortuous iliac anatomy

Weale A. R. 1, Balasubramaniam K. 2, Hardman J. 2, Horrocks M. 1

1 Department of Vascular Surgery, Royal United Hospital, Coombe Park, Bath BA1 3NG, UK; 2 Department of Radiology, Royal United Hospital, Coombe Park, Bath BA1 3NG, UK


PDF


AIM: Iliac limb occlusion following endovascular repair (EVAR) may result in limb threatening acute ischemia. The incidence is reported at up to 10% and is known to be influenced by iliac angulation and kinking of the stent graft. The aim of this study was to evaluate the performance of the Aorfix™ graft in tortuous iliac anatomy and examine the impact of the graft on the rate of iliac limb thrombosis following EVAR in a single UK centre
METHODS: We performed a retrospective review of all EVAR performed from May 1998 to May 2010. From November 2007, patients with highly angulated iliac anatomy were treated with the Aorfix™(Lombard) stent graft, or when a Zenith™(Cook) main body was chosen, the Aorfix™ iliac limbs were used with the Zenith™(Cook) device. We compared the rate of iliac limb occlusions before (group 1) and after (group 2) the adoption of this policy.
RESULTS: Two hundred and sixty seven patients underwent EVAR (group 1 n=129; group 2 n=138). In group 1, eight patients had a unilateral iliac limb occlusion (6.2%). Six of the patients had >90° iliac angulation, one had an unrecognised limb stenosis, and one patient had the stent landed in the external iliac. In group 2 there were no limb occlusions. Of the 138 patients, 25 had iliac angulation of >90°. Of these 25, eighteen were treated with the Aorfix™ stent graft system because of iliac angulation, and 7 were treated with Aorfix™ legs and Zenith™ bodies.
CONCLUSION: The rate of early iliac limb occlusion following EVAR in patients with angulated iliac anatomy can be substantially reduced by using the flexible Aorfix® stent graft system.