Monitoring depth of sedation: evaluating the agreement between the Bispectral Index, qCON and the Entropy Module’s State Entropy during flexible bronchoscopy

Johannes N. MÜLLER 1, 2, Matthias KREUZER 1, 3, 4, Paul S. GARCÍA 3, 4, Gerhard SCHNEIDER 5, Hubert HAUTMANN 2

1 Department of Anesthesiology, Rechts der Isar Hospital, Technical University of Munich, Munich, Germany; 2 Medical Clinic and Policlinic, Rechts der Isar Hospital, Technical University of Munich, Munich, Germany; 3 Department of Anesthesiology, Atlanta VA Hospital, Decatur, GA, USA; 4 Emory University Hospital, Atlanta, GA, USA; 5 Department of Anesthesiology, HELIOS University Hospital, Witten/Herdecke University, Wuppertal, Germany


PDF  


BACKGROUND: We investigated the correlation and agreement of three depth of anesthesia indices, Bispectral Index (BIS), qCON and state entropy (SE) during propofol sedation, because there is extensive literature that deals with the comparability of these indices during general anesthesia, but not during sedation.
METHODS: We recorded electroencephalogram (EEG) and SE trend data, using the Entropy Module from 21 patients who underwent elective bronchoscopy with target-controlled infusion of propofol. EEG data were replayed to BIS and qCON with an EEG player. We calculated the Spearman correlation to evaluate similarities in index trend behavior and estimated the general index agreement of displaying the same anesthetic level, i.e., wakefulness, sedation and anesthesia. We used coughing episodes and the bronchoscope placement to investigate the index for possible arousal reactions.
RESULTS: We found a high to very high correlation of the indices’ trend during the procedure. Furthermore, qCON was significantly lower than BIS and SE. All indices increased significantly after bronchoscope placement and coughing. The agreement of BIS/SE was 68% and around 50% for BIS/qCON and qCON/SE. The median duration of disagreement by >10 points was 39 s for BIS/SE and around 75 s for qCON with BIS and SE.
CONCLUSIONS: The results indicate a high probability of similar index performance between SE, BIS and qCON with the caveat of a different index scaling for qCON. These results can help the user of these monitoring devices to translate findings from one index to the other.


KEY WORDS: Consciousness monitors - Bronchoscopy - Deep sedation

inizio pagina

Publication History

Issue published online: June 7, 2017
Article first published online: February 8, 2017
Manuscript accepted: January 26, 2017
Manuscript revised: January 24, 2017
Manuscript received: February 10, 2016

Per citare questo articolo

Müller JN, Kreuzer M, García PS, Schneider G, Hautmann H. Monitoring depth of sedation: evaluating the agreement between the Bispectral Index, qCON and the Entropy Module’s State Entropy during flexible bronchoscopy. Minerva Anestesiol 2017;83:563-73. DOI: 10.23736/S0375-9393.17.11262-9

Corresponding author e-mail

hautmann@tum.de