Home > Riviste > Minerva Anestesiologica > Fascicoli precedenti > Minerva Anestesiologica 2008 July-August;74(7-8) > Minerva Anestesiologica 2008 July-August;74(7-8):375-80

ULTIMO FASCICOLO
 

ARTICLE TOOLS

Estratti

MINERVA ANESTESIOLOGICA

Rivista di Anestesia, Rianimazione, Terapia Antalgica e Terapia Intensiva


Official Journal of the Italian Society of Anesthesiology, Analgesia, Resuscitation and Intensive Care
Indexed/Abstracted in: Current Contents/Clinical Medicine, EMBASE, PubMed/MEDLINE, Science Citation Index Expanded (SciSearch), Scopus
Impact Factor 2,036


eTOC

 

ORIGINAL ARTICLES  


Minerva Anestesiologica 2008 July-August;74(7-8):375-80

lingua: Inglese

Transcutaneous carbon dioxide monitoring in spontaneously breathing, nonintubated patients in the early postoperative period

Fanelli G., Baciarello M., Squicciarini G., Malagutti G., Zasa M., Casati A.

Department of Anesthesiology, Intensive Care and Pain Medicine, University of Parma, Parma, Italy


FULL TEXT  


Background. The authors investigated the accuracy of transcutaneous capnometry (TcPCO2) in estimating arterial blood carbon dioxide partial pressure (PaCO2) during spontaneous breathing in patients admitted to our surgical intensive care unit (ICU).
Methods. Serial TcPCO2 and PaCO2 measurements were taken in stable patients undergoing postoperative monitoring after major abdominal, vascular, or thoracic surgery. Patients were enrolled ≥12 hours after extubation. Exclusion criteria were pulmonary dysfunction, hemodynamic instability, or anemia. Linear regression, mixed models, and Bland-Altman analyses were used to compare accuracy and correlation between the two variables. Data are presented as means (95% confidence intervals).
Results. PaCO2 values ranged between 26 mmHg and 52 mmHg. Mean values for TcPCO2 and PaCO2 were 35.3 (33.8-36.8) mmHg and 39.2 (37.6-40.7) (P<0.001). Bland-Altman analysis showed a bias of -3.8 (-5.1-2.5) mmHg with upper and lower limits of agreement of 5.6 (3.4-7.9) mmHg and -13.3 (-15.6 – -11.1) mmHg, respectively. Thirty-one (55.4%) measurements disagreed by 3 mmHg or more. Linear mixed model analysis with adjustment for repeated measurements showed low correlation (r=0.63; P<0.0005). There were small but significant differences between patients in the correlation coefficient (P=0.04).
Conclusion. TcPCO2 showed moderate bias with wide limits of agreement when compared to PaCO2 in our patients. The regression model showed low correlation of the measurements when adjusting for repeated measurements and between-subject variances. TcPCO2 may not provide an accurate quantitative estimation of PaCO2 in stable, spontaneously breathing, nonintubated patients in the early postoperative period.

inizio pagina

Publication History

Per citare questo articolo

Corresponding author e-mail