Advanced Search

Home > Journals > Minerva Cardioangiologica > Past Issues > Minerva Cardioangiologica 2016 October;64(5) > Minerva Cardioangiologica 2016 October;64(5):507-16

ISSUES AND ARTICLES   MOST READ   eTOC

CURRENT ISSUEMINERVA CARDIOANGIOLOGICA

A Journal on Heart and Vascular Diseases


Official Journal of the Italian Society of Angiology and Vascular Pathology
Indexed/Abstracted in: EMBASE, PubMed/MEDLINE, Science Citation Index Expanded (SciSearch), Scopus
Impact Factor 0,752

 

Minerva Cardioangiologica 2016 October;64(5):507-16

 ORIGINAL ARTICLES

Impact of atherosclerotic plaque components and their distribution on stent deployment: an intravascular-ultrasound virtual histology observational study

Manolis VAVURANAKIS, Theodore G. PAPAIOANNOU, Ourania A. KATSAROU, Dimitrios A. VRACHATIS, Elias A. SANIDAS, Gerasimos SIASOS, Konstantinos I. KALOGERAS, Dimitrios SCHIZAS, Christodoulos I. STEFANADIS, Dimitris TOUSOULIS

First Department of Cardiology, Hippokration Hospital, Medical School, National and Kapodistrian University of Athens, Athens, Greece

BACKGROUND: The aim of this study was to evaluate how the spatial distribution of each plaque element, defined by intravascular-ultrasound virtual histology (IVUS-VH), may affect stent deployment even at high inflation pressures.
METHODS: Thirty-two patients undergoing direct percutaneous coronary intervention and IVUS were evaluated. Fifty-two lesions were treated with drug-eluting stents. Pre-stenting lumen area and real (Rcssla) and average cross-sectional stent lumen area (Acssla) were measured along the whole lesion. Ideal cross-sectional stent lumen area (Icssla) was calculated. Plaque composition was characterized by IVUS-VH. The spatial distribution of each plaque element was quantified by a novel image analysis tool measuring the area and percentage of each plaque component that was adjacent to the lumen. Average stent deployment was defined as: [1 – (Icssla–Acssla)/Icssla]×100%.
RESULTS: Stent expansion was significantly less at the site of maximum calcification compared to the average stent deployment (80±9% vs. 85±13%, P=0.044, respectively). Furthermore, wherever calcium was adjacent to the lumen, stent expansion was impaired compared to sites where calcium was non-luminal (70±23% vs. 80±9%, P=0.01, respectively). In contrast, at the site of maximum necrotic core, stent deployment showed a trend to be less compromised, compared to the average stent deployment.
CONCLUSIONS: An interaction was found between plaque components and their distribution and stent deployment even at high inflation pressures.

language: English


FULL TEXT  SUPPLEMENTARY MATERIAL  REPRINTS

top of page