Home > Journals > Minerva Anestesiologica > Past Issues > Minerva Anestesiologica 2001 May;67(5) > Minerva Anestesiologica 2001 May;67(5):359-69

CURRENT ISSUE
 

ARTICLE TOOLS

Reprints

MINERVA ANESTESIOLOGICA

A Journal on Anesthesiology, Resuscitation, Analgesia and Intensive Care


Official Journal of the Italian Society of Anesthesiology, Analgesia, Resuscitation and Intensive Care
Indexed/Abstracted in: Current Contents/Clinical Medicine, EMBASE, PubMed/MEDLINE, Science Citation Index Expanded (SciSearch), Scopus
Impact Factor 2,036


eTOC

 

ORIGINAL ARTICLES  ANESTESIOLOGY


Minerva Anestesiologica 2001 May;67(5):359-69

language: Italian

Blood coagulation monitoring during liver transplantation: “Sonoclot” analysis and laboratory tests

Bindi M. L., Biancofiore G. D., Consani G., Cellai F., Cecconi N. *, Romanelli A. ***, Filipponi F. **, Mosca F. **, Amorese G., Vagelli A.

Azienda Ospedaliera Pisana - P.O. di Cisanello, Pisa I U.O. Anestesia e Rianimazione - UTI Postchirurgica e Trapianti Università degli Studi - Pisa
*U.O. Ematologia
**U.O. Chirurgia e Trapianti
***Istituto di Fisiologia Clinica, CNR, Pisa


FULL TEXT  


Background. Aim of this paper is to validate blood coagulation data obtained using the Sonoclot thromboelastographic analyser (Sienco Inc., Morrison, USA) by means of standard laboratory tests during orthotopic liver transplantation (OLT).
Methods. Design: comparative study between laboratory data and Sonoclot analysis on simultaneously collected blood samples. Setting: National Health System Liver Transplantation Center. Patients: fifty-one patients, both males and females, affected by terminal hepatic disease submitted to OLT were enrolled in the study. Data collection: simultaneous blood samples were collected during the pre-anhepatic, anhepatic and post-reperfusion phases of OLT; coagulation status was assessed by means of either standard Laboratory tests (INR, aPTT, Fibrinogen, PLT, D-Dimer) and Sonoclot analyser data (SonACT, Rate, Time to Peack, Downward Deflection); a statistical analysis was performed (Pearson’s χ2 test).
Results. A statistically significative correlation between the analysed data was found. The Sonoclot analyser was useful in identifying platelets dysfunction and was more sensitive to detect fibrinolysis.
Conclusions. The Sonoclot thromboelastographic analyser is a reliable device for monitoring coagulation during OLT.